Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 83 III: (1 year) CAL VER /5700,5725 DATE: 10-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 61 NUMBER OF LINES: 244 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= # # Source: # Fluke 80 Series III Multimeters Service Information (Preliminary, 2/98) # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # Sub Fluke 83,85,87 III (1 year) Capacitance CAL VER # Sub Fluke 83,85,87 (1 year) Frequency CAL VER /3325 # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 HEAD {DISPLAY TEST} 1.006 MESS Turn the meter on and press any push button to hold the 1.006 MESS meter in Display Test mode. 1.007 EVAL Are all the display segments illuminated? 2.001 MESS 2.002 DISP No warm-up time requirement. 2.002 DISP Ambient temperature: 18C - 28C. 2.002 DISP Relative humidity: less than 90%. 2.003 ASK- U 2.004 HEAD {ROTARY SWITCH TESTS} 2.005 DISP Hold down the RANGE button while turning the UUT rotary 2.005 DISP function switch from OFF to VDC. Release the RANGE 2.005 DISP button and turn the switch to {VAC}. 2.006 MATH MEM1 = 0 2.007 MEMI Enter UUT reading: 2.008 MEME 2.009 MEMC display 12U 3.001 DISP Rotate the UUT function switch to {VDC}. 3.002 MATH MEM1 = -32 3.003 MEMI Enter UUT reading: 3.004 MEME 3.005 MEMC display 12U 4.001 DISP Rotate the UUT function switch to {mVDC}. 4.002 MATH MEM1 = -64 4.003 MEMI Enter UUT reading: 4.004 MEME 4.005 MEMC display 12U 5.001 DISP Rotate the UUT function switch to {Ohms}. 5.002 MATH MEM1 = -96 5.003 MEMI Enter UUT reading: 5.004 MEME 5.005 MEMC display 12U 6.001 DISP Rotate the UUT function switch to {Diode}. 6.002 MATH MEM1 = -128 6.003 MEMI Enter UUT reading: 6.004 MEME 6.005 MEMC display 12U 7.001 DISP Rotate the UUT function switch to {mA/A}. 7.002 MATH MEM1 = -160 7.003 MEMI Enter UUT reading: 7.004 MEME 7.005 MEMC display 12U 8.001 DISP Rotate the UUT function switch to {uA}. 8.002 MATH MEM1 = -192 8.003 MEMI Enter UUT reading: 8.004 MEME 8.005 MEMC display 12U 9.001 ASK+ U 9.002 HEAD {AC VOLTAGE TESTS} 9.003 DISP Rotate the UUT function switch to OFF then VAC. 9.004 DISP Connect the 5700A and the UUT as follows. 9.004 DISP [32] 5700A OUTPUT HI to UUT V Ohm Diode 9.004 DISP [32] 5700A OUTPUT LO to UUT COM 9.004 DISP [32] 5700A SENSE HI to UUT V Ohm Diode 9.004 DISP [32] 5700A SENSE LO to UUT COM 9.005 HEAD AC VOLTAGE TESTS: {400mV Range} 9.006 5700 400 350.0mV 2.2U 60H 2W 10.001 5700 400 350.0mV 3.9U 1kH 2W 11.001 5700 400 350.0mV 7.4U 5kH 2W 12.001 HEAD AC VOLTAGE TESTS: {4V Range} 12.002 5700 4 3.500V 0.020U 60H 2W 13.001 5700 4 3.500V 0.039U 1kH 2W 14.001 5700 4 3.500V 0.074U 5kH 2W 15.001 HEAD AC VOLTAGE TESTS: {40V Range} 15.002 5700 40 35.00V 0.20U 60H 2W 16.001 5700 40 35.00V 0.39U 1kH 2W 17.001 5700 40 35.00V 0.74U 5kH 2W 18.001 HEAD AC VOLTAGE TESTS: {400V Range} 18.002 5700 400 350.0V 2.0U 60H 2W 19.001 5700 400 350.0V 3.9U 1kH 2W 20.001 5700 400 350.0V 7.4U 5kH B1 2W 21.001 HEAD AC VOLTAGE TESTS: {1000V Range} 21.002 5700 1000 900V 7U 60H 2W 22.001 5700 1000 900V 13U 1kH 2W # Better TURs for the frequency tests can be achieved using a HP 3325B # function generator. If you wish to use the function generator in lieu # of the 5700A delete the 5700A based frequency tests and uncomment the # CALL statement at the bottom of this procedure. 23.001 HEAD {FREQUENCY TESTS} 23.002 DISP Press the RANGE button to lock the 400mV Range. 23.002 DISP Press the Hz button to choose the Frequency mode. 23.003 HEAD FREQUENCY TESTS: {19.999kHz Range} 23.004 5700 20 19.000kH 0.002U 150mV 2W #! WARNING: Test Tol 2, Sys Tol 1.9, TUR 1.0526 (< 4.00). 24.001 HEAD FREQUENCY TESTS: {199.99kHz Range} 24.002 5700 200 190.00kH 0.02U 150mV 2W #! WARNING: Test Tol 20, Sys Tol 19, TUR 1.0526 (< 4.00). 25.001 HEAD {FREQUENCY SENSITIVITY AND TRIGGER LEVEL} 25.002 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V AC Range} 25.003 DISP Press the RANGE button to lock the 4V Range. 25.004 5700 1000.0H 0.2U 300mV 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 26.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V DC Range} 26.002 DISP Rotate the UUT function switch to VDC. 26.002 DISP Press the RANGE button to lock the 4V Range. 26.002 DISP Press the Hz button to choose the Frequency mode. 26.003 5700 1000.0H 0.2U 1.7V 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 27.001 5700 1000.0H 1.0V S 2W 27.002 EVAL Is the UUT {displaying all zeros}? 28.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {40V DC Range} 28.002 DISP Press the RANGE button to lock the 40V Range. 28.003 5700 1000.0H 0.2U 6.0V 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 29.001 5700 1000.0H 2.0V S 2W 29.002 EVAL Is the UUT {displaying all zeros}? 30.001 HEAD {DC VOLTAGE TESTS} 30.002 DISP Press the Hz button twice to select VDC. 30.002 DISP 30.002 DISP Press the RANGE for 2s to select autorange. 30.003 HEAD DC VOLTAGE TESTS: {4V Range} 30.004 5700 4 3.500V 0.005U 2W 31.001 HEAD DC VOLTAGE TESTS: {40V Range} 31.002 5700 40 35.00V 0.05U 2W 32.001 5700 40 -35.00V 0.05U 2W 33.001 HEAD DC VOLTAGE TESTS: {400V Range} 33.002 5700 400 350.0V 0.5U 2W 34.001 HEAD DC VOLTAGE TESTS: {1000V Range} 34.002 5700 1000 1000V 2U 2W 35.001 HEAD {DC MILLIVOLT TEST} 35.002 DISP Rotate the UUT function switch to mVDC. 35.003 HEAD DC MILLIVOLT TEST: {400mV Range} 35.004 5700 400 350.0mV 1.2U 2W 36.001 HEAD {RESISTANCE TESTS} 36.002 DISP Rotate the UUT function switch to Ohms. 36.003 HEAD RESISTANCE TESTS: {400 Ohm Range} 36.004 5700 400 190.0Z 1.0U CW 37.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 37.002 5700 40 19.00kZ 0.09U CW 38.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 38.002 5700 4 1.900MZ 0.014U 2W 39.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 39.002 5700 40 19.00MZ 0.22U 2W 40.001 HEAD {CONDUCTANCE TEST} 40.002 DISP Press the Range button twice for 40nS range. 40.003 HEAD CONDUCTANCE TEST: {40nS Range} 40.004 5700 40 10.00nY 0.20U 2W 41.001 HEAD {DIODE TEST} 41.002 DISP Rotate the UUT function switch to Diode. 41.003 5700 3.000V 0.061U 2W 42.001 CALL Sub Fluke 83,85,87 III (1 year) Capacitance CAL VER 42.002 JMP 43.001 42.003 EVAL dummy 43.001 HEAD {DC MILLIAMP TESTS} 43.002 DISP Rotate the UUT function switch to mA/A. 43.003 DISP Connect the 5700A and the UUT as follows. 43.003 DISP [32] 5700A OUTPUT HI to UUT mA/uA 43.003 DISP [32] 5700A OUTPUT LO to UUT COM 43.004 HEAD DC MILLIAMP TESTS: {40mA Range} 43.005 5700 40 35.00mA 0.18U 2W 44.001 HEAD DC MILLIAMP TESTS: {400mA Range} 44.002 5700 400 350.0mA 1.6U 2W 45.001 HEAD {AC MILLIAMP TESTS} 45.002 DISP Press the BLUE button to select AC. 45.003 HEAD AC MILLIAMP TESTS: {40mA Range} 45.004 5700 40 35.00mA 0.44U 60H 2W 46.001 5700 40 35.00mA 0.44U 1kH 2W 47.001 HEAD AC MILLIAMP TESTS: {400mA Range} 47.002 5700 400 350.0mA 4.4U 60H 2W 48.001 5700 400 350.0mA 4.4U 1kH 2W 49.001 HEAD {DC MICROAMP TESTS} 49.002 DISP Rotate the UUT function switch to uA. 49.003 HEAD DC MICROAMP TESTS: {400uA Range} 49.004 5700 400 350.0uA 1.8U 2W 50.001 HEAD DC MICROAMP TESTS: {4000uA Range} 50.002 5700 4000 3500uA 16U 2W 51.001 HEAD {AC MICROAMP TESTS} 51.002 DISP Press the BLUE button to select AC. 51.003 HEAD AC MICROAMP TESTS: {400uA Range} 51.004 5700 400 350.0uA 4.4U 60H 2W 52.001 5700 400 350.0uA 4.4U 1kH 2W 53.001 HEAD AC MICROAMP TESTS: {4000uA Range} 53.002 5700 4000 3500uA 44U 60H 2W 54.001 5700 4000 3500uA 44U 1kH 2W 55.001 DISP Disconnect the 5700A from the UUT. 55.002 HEAD {DC AMP TESTS} 55.003 DISP Rotate the UUT function switch to mA/A. 55.004 DISP Connect the 5725A and UUT as follows: 55.004 DISP [32] 5725A CURRENT OUTPUT HI to UUT A 55.004 DISP [32] 5725A CURRENT OUTPUT LO to UUT COM 55.005 HEAD DC AMP TESTS: {4000mA Range} 55.006 5700 4000 3500mA 18U B1 2W 56.001 HEAD DC AMP TESTS: {10A Range} 56.002 5700 10 10.00A 0.06U B1 2W 57.001 HEAD {AC AMP TESTS} 57.002 DISP Press the BLUE button to select AC. 57.003 HEAD AC AMP TESTS: {4000mA Range} 57.004 5700 4000 3500mA 44U 60H B1 2W 58.001 5700 4000 3500mA 44U 1kH B1 2W 59.001 HEAD AC AMP TESTS: {10A Range} 59.002 5700 10 10.00A 0.14U 60H B1 2W 60.001 5700 10 10.00A 0.14U 1kH B1 2W #61.001 CALL Sub Fluke 83,85,87 (1 year) Frequency CAL VER /3325 61.001 END #! T.U.R.s less than 4.00: 5 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 7 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.